Please refer to your institution’s standard operating procedures for cleaning surfaces contaminated with blood or bodily fluids.

For general cleaning of MAXAIR® Helmets, Batteries, Chargers, Liners, Helmet Covers, and Power Cords, we recommend using a clean, damp cloth with a mild, skin-friendly soap.

For high-level cleaning, MAXAIR® recommends two alternatives:

  • PDI Super Sani-Cloth® Germicidal Disposable Wipe, Q55172

  • Clorox Healthcare® Hydrogen Peroxide Cleaner Disinfectant, CLO30825

All MAXAIR® Face/Head Covers are single-use items and are not designed for cleaning or reuse.

MAXAIR® Filters and Pre-Filters should be discarded as contaminated waste if the Helmet’s yellow LED remains illuminated for more than approximately 10 seconds after the Helmet is powered on. They should also be discarded whenever they are soiled, visibly damaged, or contaminated with bodily fluids.

WARNING

Before using any cleaning agent on any material, we recommend testing the agent on a sample to evaluate both short- and long-term effects and help ensure product and user safety. MAXAIR® Systems is not responsible for the results of cleaning procedures that fall outside the cleaning agents and protocols identified in this Technical Bulletin or outside the cleaning agent manufacturer’s product-use and safety recommendations for the key materials listed herein.

Refer to Technical Bulletin MAXAIR® Systems Cleaning, P/N 04123046, for additional details.